PXI-2532B
Description
Model Explanation
PXI: PXI bus platform (modular instrument standard for testing and measurement)
2532: Product serial number, representing a 512 cross-point high-density matrix switch
Alias: PXI-2532, NI-2532, 778572-32 (part number)
Function Positioning: Reconfigurable matrix switch, supporting multiple topologies such as single-line and double-line, used for high-precision signal switching by NI
Technical Parameters
Cross-point Number: 512 (composed of 16 groups of 2×16 matrices) NI
Topology Configuration (Single Line): 4×128, 8×64, 16×32, dual 4×64, dual 8×32, dual 16×16, four 4×32 NI
Topology Configuration (Double Line): 4×64, 8×32, 16×16, dual 4×32 NI
Relay Type: Reed (dry contact), rhodium contacts NI
Maximum Switching Voltage: 300 Vrms (CAT II)
Maximum Switching Current: 0.5 A/channel, maximum switching power 10 W
Bandwidth: Single line ≥ 30 MHz, double line ≥ 25 MHz
Thermal Offset: < 1 μV (ultra-low thermoelectric potential)
Isolation Level: 150 V CAT II
Trigger Characteristics: PXI trigger line 0–7, minimum pulse width 150 ns; output trigger 1–62 μs, software selectable NI
Size: 3U PXI module (160 × 100 × 30 mm)
Weight: Approximately 0.8 kg
Status:停产; recommended replacement: PXIe-2532BNI
Interface and Communication Configuration
PXI Bus Interface: Standard 3U PXI backplane connector, providing power supply, trigger, clock and communication NI
Front-end I/O Connector: 2 × 160-pin Samtec BTE-EM series high-voltage connector NI
Terminal Block (optional): SCB-2586, etc., supports ribbon cable or screw terminal leads
Control Communication: PXI bus, compatible with NI- SWITCH, MAX configuration software NI
Trigger Synchronization: Hardware trigger input/output, supports synchronization with other instruments in the PXI system NI
Expansion Capability: Multiple modules can be cascaded through the terminal block to expand the matrix size NI Core function
High-density matrix switching: 512 cross points per slot, flexible configuration of multiple topologies, meeting the requirements of multi-channel signal routing for NI.
Ultra-low thermal offset: < 1 μV, suitable for low-level, high-precision measurements (such as thermocouple, sensor signals).
High-voltage isolation: 150 V CAT II isolation, allowing safe switching of mixed high-low voltage signals.
Hardware triggering and synchronization: nanosecond-level triggering, supporting high-speed scanning and multi-instrument synchronization, enhancing test throughput for NI.
Relay life monitoring: onboard relay count, predicting maintenance cycles, reducing unexpected downtime for NI.
Reconfigurable topology: flexible switching of single-line/double-line matrix through software and terminal block, adapting to different testing requirements for NI. Applicable scenarios
Aerospace electronic automation testing: Multi-channel signal switching and fault injection.
Semiconductor wafer testing: High-density probe card signal routing.
Automotive ECU verification: Sensor, load signal simulation and switching.
Industrial sensor network monitoring: Multi-node round-trip collection and switching.
Precise instrument calibration: High-precision switching of low-level and weak signals.
User Manual
Usage Tips
Installation: Insert into the standard 3U PXI chassis slot, lock the panel, and ensure the backplane connector makes reliable contact.
Configuration: Select the matrix topology through NI-SWITCH or MAX software, and define the channel mapping and trigger parameters NI.
Wiring: Connect the 160-pin connector to the terminal block (such as SCB-2586), and then transfer it to the tested device; for high voltage / large current, connect according to specifications to avoid overload.
Operation Monitoring: Use the onboard relay counting function to regularly read the number of switches, and evaluate the lifespan NI.
Environment: Working temperature 0–50 °C, avoid strong electromagnetic interference, dust and condensation.
Maintenance Instructions
Regular Inspection: Check the connector tightness, and whether the terminal block wiring is loose every quarter. Clean the surface dust.
Relay Life Management: Typical lifespan of reed relays: 10 V/100 mA approximately 10⁷ times; 20 V/500 mA approximately 5×10⁶ times; 100 V/10 mA approximately 5×10⁵ times. Replace the module in time when approaching the lifespan.
Model Interpretation
PXI: PXI bus platform (modular instrument standard for testing and measurement)
2532: Product serial number, representing 512 cross-point high-density matrix switch
B: Hardware version upgrade, optimizing relay lifespan, thermal electromotive force and bandwidth performance
Alias: NI-2532B, PXI-2532B (PXI bus version)
Function Position: Reconfigurable matrix switch, supporting single-line / double-line topology, suitable for high-precision signal switching
Technical Parameters
Cross-point Number: 512 (16×32 matrix architecture) NI
Topology Configuration (Single Line): 4×128, 8×64, 16×32, dual 4×64, dual 8×32, dual 16×16, four 4×32
Topology Configuration (Double Line): 4×64, 8×32, 16×16, dual 4×32
Relay Type: Reed (Reed), Rhodium contact, mechanical life 1×10⁹ times
Maximum Switching Voltage: 100 VDC / 100 Vp (CAT I)
Maximum Switching Current: 0.5 A / channel, maximum switching power 10 W
Bandwidth (–3 dB, 50 Ω): Single line ≥ 30 MHz, double line ≥ 25 MHz
Thermal Electromotive Force (Thermal EMF): Single line < 50 μV, double line < 20 μV NI
Channel Isolation (1 MHz): > 44 dB
Switching Speed: Maximum 2000 times / second NI
Trigger Characteristics: PXI trigger line 0–7, minimum pulse width 150 ns
I/O Connector: 2×150-pin Samtec ERM8-EM series
Size: 3U PXI module (160×100×30 mm)
Weight: Approximately 0.8 kg
Working Temperature: 0–50 °C
Interface and Communication Configuration
PXI Bus Interface: Standard 3U PXI hybrid backplane connector, providing power supply, trigger, clock and communication NI
Front-end I/O Connector: 2×150-pin Samtec ERM8-EM high-voltage connector, supporting high-density signal output Terminal Rack (Optional): TB-2640B, TB-2641B, etc., supports screw terminals or ribbon cable connections to NI
Control Communication: PXI bus, compatible with NI-SWITCH, MAX, LabVIEW, TestStand software from NI
Trigger Synchronization: Hardware trigger input/output, supports synchronization with other instruments in the PXI system, enhancing test throughput
Expansion Capability: Multiple modules can be cascaded with R1616 cables through the TB-264xB terminal rack, expandable to a 4×256 matrix Core function
High-density matrix switching: 512 cross points per slot, supports 10+ topologies, meets the requirements of multi-channel signal routing for NI.
Ultra-low thermal offset: less than 20 μV for dual wires, suitable for low-level high-precision measurement of thermocouples, sensors, etc. NI.
High-voltage isolation and protection: 100 V CAT I isolation, can safely switch mixed high and low voltage signals.
High-speed switching and synchronization: 2000 switching cycles per second, hardware-triggered nanosecond-level synchronization, improves test efficiency NI.
Relay life monitoring: onboard relay counter, predicts maintenance cycle, reduces unexpected downtime NI.
Flexible topology reconfiguration: software and terminal block cooperation, quickly switch between single-line and dual-line matrices, adaptable to different testing requirements NI. Applicable scenarios
Aerospace electronic automation testing: multi-channel signal switching, fault injection and aviation bus testing.
Semiconductor wafer testing: high-density probe card signal routing, parameter testing and aging screening.
Automotive ECU verification: sensor / load signal simulation, wiring harness conduction test and fault diagnosis.
Industrial sensor network monitoring: multi-node round-trip acquisition, signal conditioning and data recording.
Precision instrument calibration: high-precision switching and calibration system setup for low-level and weak signals.
Military and research testing: high-reliability, high-density signal routing and execution of complex test sequences.
User and maintenance instructions
Key usage points
Installation: Insert into the standard 3U PXI chassis slot, lock the panel, ensure reliable contact of the backplane connectors.
Configuration: Select the matrix topology through NI-SWITCH or MAX software, define channel mapping and trigger parameters NI.
Cabling: Connect the front 150-pin connector to the TB-2640B/TB-2641B terminal block, then transfer to the tested equipment; for high voltage / large current, connect according to specifications to avoid overload NI.
Operation monitoring: Use the onboard relay counting function to regularly read the switching times, evaluate the lifespan NI.
Environment: Working temperature 0–50 °C, avoid strong electromagnetic interference, dust and condensation.
Maintenance instructions
Regular inspection: Check the connector tightness, whether the terminal block wiring is loose, and clean the surface dust every quarter.
Relay life management: Typical life: Low power (10 V/100 mA) approximately 10⁷ times; High power (50 V/0.5 A) approximately 5×10⁵ times, replace the module in time when approaching the lifespan.
Fault troubleshooting: When the channel is not connected or has poor contact, first check the wiring and terminal block, then diagnose the relay status through software; check the grounding and shielding when the thermal offset exceeds the limit.
Storage: When not in use for a long time, disconnect the power supply, place it in a dry, dust-proof environment, and avoid the connectors getting wet and oxidized.
Get a Quote