PXI-4072
Description
II. Model Interpretation
PXI: 3U PXI standard bus, single slot, compatible with PXI/PXIe hybrid chassis.
4072: FlexDMM series, representing 6½ digit, ±300 V, with LCR, 1.8 MS/s isolation digital instrument.
778270-01: Original part number, used for ordering and traceability.
Positioning: DMM + LCR + digital instrument in one unit, replacing bench 6½ digit meter + LCR meter + isolated acquisition card, suitable for medium voltage, high precision + impedance testing requirements.
III. Technical Parameters (25 ℃ ± 1 ℃, calibration cycle 2 years, preheating 30 minutes)
(1) DMM Function
DC Voltage (DCV)
Range: 100 mV / 1 V / 10 V / 100 V / 300 V
2-year accuracy: 15 ppm reading + 0.5 ppm (10 V)
AC Voltage (True RMS)
Range: 100 mV - 300 V
Bandwidth: 10 Hz - 100 kHz
Accuracy: 0.02% reading + 50 μV (10 V @ 1 kHz)
DC Current (DCI)
Range: 10 mA / 100 mA / 1 A
Accuracy: 0.05% reading + 20 nA (1 A)
AC Current (True RMS)
Range: 10 mA - 1 A
Bandwidth: 10 Hz - 100 kHz
Resistance (Ω)
2-wire: 10 Ω - 5 GΩ; 4-wire: 10 Ω - 1 MΩ
4-wire accuracy: 0.01% reading + 10 mΩ (1 kΩ)
Frequency / Period: 10 Hz - 100 kHz, accuracy 0.01%
Resolution / Rate: 6½ digit (22 bit); 6½ digit 100 times / second, 5½ digit 3,000 times / second
(2) LCR Measurement (Core Feature)
Capacitance C: 300 pF - 10,000 μF; basic accuracy 0.25%; rate up to 20 times / second
Inductance L: 1 mH - 5 H; basic accuracy 0.25%; rate up to 20 times / second
Measurement Method: 2-wire AC current method, supports open circuit / short circuit compensation NI
(3) Isolated Digital Instrument
Sampling Rate: Up to 1.8 MS/s (single channel)
Resolution: 10 - 22 bit adjustable
Input Range: ±10 V / ±100 V / ±300 V (isolated)
Bandwidth: DC - 500 kHz
Isolation Voltage: 300 Vrms (channel - ground)
Trigger: Edge / Window / Pulse Width / Digital Trigger, delay ≤ 2 μs
(4) Electrical / Mechanical / Environmental
Power Supply: PXI backplane; typical 8 W, maximum 12 WNI
Connectors: 5× safety banana sockets (gold plated low thermoelectric potential) + 9-pin DIN (trigger / AUX I/O)
Size: 3U single slot, 2.0 cm × 13.0 cm × 21.6 cmNI
Weight: Approximately 370 gNI
Safety: IEC 61010-1, CAT III 300 V
Operating Temperature: 0 ℃ - 55 ℃ (accuracy guaranteed 15 ℃ - 35 ℃)
Calibration Cycle: 2 years (NIST traceable)
IV. Interface and Communication Configuration
(1) Front-End I/O
5× safety banana sockets (gold-plated low thermoelectric potential)
HI/LO: Voltage / Current / Resistance Main Terminal
HI Sense/LO Sense: 4-wire Resistance Sensing Terminal
Ground: Digital Instrument Isolation Ground
9-pin DIN (AUX I/O): Trigger Input/Output, PXI Trigger Synchronization, SCXI Communication
(II) Bus and Software
Bus: Standard PXI, Compatible with PXIe Mixed Slot
Driver: NI-DMM (IVI) + NI-Digitizer
Software: LabVIEW, TestStand, LabWindows/CVI, Python, C/C++, .NET
Configuration: NI MAX (Identification, Self-Check, Self-Calibration, Trigger Settings)
Synchronization: PXI Trigger, PFI, Multi-module Synchronization Delay ≤ 2 μs
VII. Core Functions
6½-digit High Precision DMM: DC Accuracy ppm Level, Suitable for Precise Parameter Testing.
±300 V High Voltage Isolation: 300 Vrms Isolation, Safe Measurement of Medium High Voltage Grounded Signals.
LCR Function: Capacitance / Inductance 0.25% Basic Accuracy, Covers Production Test of Components.
1.8 MS/s Isolated Digital Converter: Captures 300 V Level Transients, Ripples, Switching Waveforms.
4-wire Kelvin Resistance: Eliminates Lead Error, Suitable for Precise Resistors / RTD.
True RMS AC Measurement: Accurate Measurement of Frequency Converters, Switching Power Supplies, etc. Non-sinusoidal Waveforms.
Three-in-One Integration: Single Slot Replacement of DMM + LCR + Isolated Acquisition Card, Simplifies System, Reduces Cost.
Dual Internal Calibration: Internal Self-Calibration to Compensate for Temperature Drift, External Calibration to Ensure Long-Term Precision.
VI. Application Scenarios
Electronic Component Mass Production Testing: Capacitors, Inductors, Resistors, Diodes' Parameters + Impedance Testing.
Medium High Voltage Power Supply / Frequency Converter Testing: 300 V Level Power Supply, Inverters, Charging Module's Static Parameters + Ripple / Transient Waveforms.
Industrial Automation ATE: Consumer Electronics, Household Appliances, Automotive Electronics' Batch Precise Testing.
Precise Sensor Calibration: RTD, Thermocouple, Pressure Sensor's Resistance / Voltage + Dynamic Response Testing.
Power Electronics Research and Development: MOSFET, IGBT, etc. Device's Static Parameters + Switching Waveform Characterization.
General Laboratory Precision Measurement: Covers Voltage / Current / Resistance / LCR / Waveform Acquisition Requirements.
VII. Usage and Maintenance Instructions
(I) Installation and Power On
Disconnect power, insert PXI single slot, tighten panel screws.
Power On Preheat for 30 minutes to ensure 6½-digit accuracy.
NI MAX confirms equipment is normal, self-checks without errors.
(II) Wiring and Operation
High Voltage Safety: Use safety plugs ≥ 100 V; Do not input > 300 V.
4-wire Resistance: Current Terminal (HI/LO) and Sensing Terminal (Sense) are wired separately.
LCR Measurement: Use Open Circuit / Short Circuit Compensation to Eliminate Parasitic Resistance Error NI.
Current Range Protection: Replace 1 A Fast Fuse after Overload Power Off.
(III) Maintenance and Calibration
Regular Calibration: External Traceable Calibration Every 2 Years; Daily Internal Self-Calibration to Compensate for Temperature Drift.
Cleaning: Use Dry Cloth / Soft Brush to Remove Dust, Keep Interfaces Dry, Cover with Dust Cap When Not in Use for a Long Time.
Environment: 0 ℃~55 ℃, No Condensation, Away from Strong Vibration / Strong Electromagnetic Interference.
Prohibited: Do Not Use Beyond Measured Range / Beyond Isolation Voltage; Avoid Large Current Impact.
Get a Quote