DU-2/MKII
May 12, 2026

DU-2/MKII

Description

1. Product Overview

DU-2/MKII is the MKII upgraded second-generation divided-potential electrochemical trace oxygen sensor.
Optimized with improved electrolyte formula and enhanced anti-poisoning structure, it is specially designed for
ppm-level ultra-low oxygen concentration detection. It features lower zero drift, faster response, superior
temperature adaptability and long-term operational stability. Adopting standard DU concentric ring interface,
it is fully compatible with DPD series mounting bases, DIM5-1 signal conversion modules and
DELOMATIC-3/DGU1 control units. Widely applied in industrial inert gas monitoring, semiconductor fabrication,
laboratory glove boxes and professional trace oxygen analysis instruments.

2. Key Features

  • MKII second-generation upgraded version with optimized internal formula

  • High precision ppm-level ultra-low oxygen measurement capability

  • Fast T90 response time and rapid signal recovery

  • Ultra-low monthly drift, excellent long-term stability

  • Low temperature coefficient, stable output under variable temperature

  • Enhanced anti-interference and anti-poisoning performance against background mixed gases

  • Standard DU concentric ring interface, plug-and-play replacement

  • Sealed compact cylindrical structure, easy installation and field maintenance

  • Long service life, suitable for 24-hour continuous online monitoring

3. Technical Specifications

3.1 Performance Parameters

ParameterSpecification
Product ModelDU-2/MKII
Working PrincipleDivided-potential Electrochemical
Measuring Range0~10 ppm (custom range available)
T90 Response Time≤ 7 s
Full Scale Accuracy±2% F.S.
Linearity Error≤ ±2% F.S.
Monthly Long-term Drift≤ 0.8% F.S.
Temperature Coefficient≤ 2.8% / °C
Operating Temperature0°C ~ +50°C
Operating Humidity0%~95% RH (Non-condensing)
Operating Pressure80 kPa ~ 120 kPa
Design Service Life≥ 24 months
Shelf Life12 months (sealed original packaging)

3.2 Mechanical & Interface Parameters

ParameterSpecification
Package TypeStandard cylindrical DU package
Interface ModeDU standard gold-plated concentric ring
Housing MaterialHDPE High Density Polyethylene
Net WeightApprox. 5 g
Matching BaseDPD02027 / DPD08063
Matching ModuleDIM5-1 / DELOMATIC-3/DGU1

4. Application Scope

  • Trace oxygen monitoring for nitrogen, argon and hydrogen production industry

  • Semiconductor process gas and protective atmosphere detection

  • Laboratory glove boxes, inert atmosphere furnaces and vacuum cabinets

  • Portable trace oxygen analyzers and fixed online monitoring instruments

  • Petrochemical and metallurgical shielding gas pipeline monitoring

  • Residual oxygen detection for food and pharmaceutical packaging systems

5. Installation and Operation Instructions

5.1 Installation Requirements

  1. Install vertically with gas inlet upward; keep the air inlet unobstructed.

  2. Use matched DPD dedicated base; insert vertically in place without forced twisting.

  3. Adopt sealed gas path connection; PTFE tubing is recommended.

  4. Recommended gas flow rate: 200~500 mL/min; avoid excessive airflow impact.

  5. Keep away from high temperature heat sources, strong vibration and corrosive gas environments.

5.2 Warm-up and Calibration

  1. Power on and warm up for 10–15 minutes before calibration and formal measurement.

  2. Zero Calibration: Apply high-purity nitrogen (0 ppm O₂) to complete zero point calibration.

  3. Span Calibration: Use standard low-oxygen calibration gas for span adjustment.

  4. Recalibration is required when ambient temperature or working conditions change greatly.

6. Safety Precautions

  1. This sensor is non-explosion-proof; do not use in flammable and explosive hazardous areas.

  2. Prevent condensed water, oil mist and dust from entering the air inlet.

  3. Take ESD protection measures during handling and installation.

  4. Do not disassemble the sensor housing privately to avoid electrolyte leakage and permanent failure.

  5. Avoid long-term exposure to strong corrosive gases such as SO₂, Cl₂ and H₂S.

7. Maintenance and Storage

  1. Perform zero and span calibration regularly every 3–6 months according to service conditions.

  2. Keep the air inlet filter clean; replace the filter once blocked or contaminated.

  3. Store in sealed original package at 10~30°C, 30%~70% RH in a cool and dry place.

  4. Avoid direct sunlight, high temperature and humid stacked storage.

8. Troubleshooting

Fault PhenomenonPossible CauseSolution
No signal outputPoor contact / wrong insertion directionReinstall sensor correctly; check base contact
Unstable readingInsufficient warm-up / gas flow fluctuationFully preheat; stabilize gas flow rate
Slow response speedAir inlet blocked / sensor agingClean filter; recalibrate or replace sensor
Large measurement driftTemperature mutation / expired calibrationTemperature compensation and recalibration

9. Warranty & Disclaimer

  • Model: DU-2/MKII Trace Oxygen Sensor

  • Warranty period: 12 months from delivery date under normal installation and use.

  • Warranty shall be invalid for damage caused by forced disassembly, improper installation, liquid immersion and artificial damage.

  • All specifications are subject to technical upgrade without prior notice.

  • The manufacturer reserves the right of final interpretation of product structure and performance.