DELOMATIC-3/DGU1
May 12, 2026

DELOMATIC-3/DGU1

Description


1. Product Overview

DELOMATIC-3/DGU1 is a high-precision trace oxygen detection control module, specially designed for online monitoring and signal conversion of trace oxygen in industrial gas.
It integrates oxygen sensor signal acquisition, temperature compensation, linear algorithm processing, digital output and analog output into one module. Featuring high stability, strong anti-interference, compact structure and plug-and-play design, it is perfectly matched with AOP / CM / CRM series oxygen sensors, widely used in nitrogen making, argon making, hydrogen industry, semiconductor, laboratory glove box and gas analysis instruments.

2. Key Features

  • Model: DELOMATIC-3/DGU1 Trace Oxygen Transmitter Module

  • Built-in high-precision temperature compensation algorithm

  • Support matching mainstream electrochemical trace oxygen sensors

  • Stable linear output, automatic calibration function

  • Dual signal output: analog + digital signal

  • Industrial grade anti-interference design, suitable for complex on-site environment

  • Compact size, easy embedded installation and system integration

  • Low power consumption, long-term continuous operation without drift

  • Simple calibration, no need for professional debugging

3. Technical Specifications

3.1 Electrical Parameters

ParameterSpecification
ModelDELOMATIC-3/DGU1
Supply VoltageDC 12V ~ 24V
Working Current≤ 80mA
Measuring Range0~10ppm / 0~100ppm (optional)
Detection Accuracy±2% FS
Resolution0.01ppm
Warm-up Time≤ 5 minutes
Analog Output4~20mA
Digital CommunicationRS485, Modbus-RTU Protocol
Temperature CompensationAutomatic full-range compensation

3.2 Environmental Parameters

ParameterSpecification
Operating Temperature0℃ ~ +50℃
Working Humidity10%~95% RH, non-condensing
Storage Temperature-20℃ ~ +60℃
Working Pressure80~120kPa
Protection LevelIP40

3.3 Mechanical Structure

ParameterSpecification
Installation MethodDIN rail / screw fixed mounting
Outline DimensionCompact modular size
Shell MaterialEngineering plastic
Net WeightApprox. 45g

4. Application Scope

  • Industrial nitrogen, argon, hydrogen trace oxygen online monitoring

  • Semiconductor manufacturing process gas detection

  • Laboratory glove box and inert atmosphere equipment

  • Portable and fixed trace oxygen analyzer supporting module

  • Food and pharmaceutical packaging residual oxygen monitoring

  • Petrochemical and metallurgical protective gas monitoring system

5. Pin Definition & Wiring Instruction

  1. Power Terminal

    • VCC: DC 12~24V positive

    • GND: Power negative and signal common ground

  2. Sensor Interface

    • Match dedicated sensor socket, support direct plug of DGU series sensor

  3. Signal Output

    • 4~20mA: Analog signal output

    • A / B: RS485 communication line

6. Installation & Operation Guide

6.1 Installation

  1. Adopt horizontal installation, keep the module away from heat source and corrosive gas.

  2. Wiring strictly follow pin definition, avoid reverse connection of power supply.

  3. Sensor wiring should be shielded wire, separate from power cable to avoid interference.

  4. Keep good ventilation around the module, do not cover or seal it completely.

6.2 Calibration Operation

  1. Power on the module and preheat for 5 minutes before calibration.

  2. Zero Calibration: Access high-purity nitrogen (0ppm), complete zero point setting.

  3. Span Calibration: Access standard gas of known oxygen concentration for span calibration.

  4. After calibration, the module automatically saves parameters and enters normal measurement state.

7. Safety Precaution

  • Do not use overvoltage power supply to avoid module burnout.

  • Avoid working in strong corrosion, flammable and explosive gas environment.

  • Do not disassemble the module privately, otherwise the warranty will be invalid.

  • Do not let liquid, dust enter the shell and sensor interface.

  • During thunderstorm season, do good grounding and lightning protection measures.

8. Maintenance

  • Regularly check wiring looseness and terminal oxidation every 3 months.

  • Recalibrate every 3~6 months according to on-site environment.

  • Keep the module surface clean and dry, avoid oil and dust accumulation.

  • Long-term shutdown should cut off power and store in dry environment.

9. Troubleshooting

Fault PhenomenonCauseSolution
No display & no powerPower supply abnormal / fuse damagedCheck input voltage and power wiring
Measurement value drifting greatlyUnfinished warm-up / sensor agingFully preheat; recalibrate or replace sensor
No 4-20mA outputWiring error / output channel faultCheck analog output wiring
RS485 communication failureA/B reverse connection / address conflictSwap A/B line; modify communication address
Unstable readingNearby electromagnetic interferenceUse shielded wire and reliable grounding

10. Warranty & Disclaimer

  • Product Model: DELOMATIC-3/DGU1

  • Warranty Period: 12 months from the date of delivery under normal use

  • The warranty does not cover damage caused by unauthorized disassembly, wrong wiring, overvoltage and man-made damage.

  • All specifications are subject to final product standard, and the design and parameters can be updated without prior notice.

  • The manufacturer reserves the right of final interpretation of product performance and parameters.