CM-2
May 12, 2026

CM-2

Description

1. Product Overview

The CM-2 is a compact, high-precision electrochemical trace oxygen sensor designed by ASAIR (Guangzhou Aosong Electronics Co., Ltd.) for accurate detection of ultra-low oxygen concentrations in complex gas environments. It features a miniaturized structure, low power consumption, fast response, and excellent long-term stability, making it ideal for integration into portable analyzers, medical devices, and industrial gas monitoring systems that require strict trace oxygen measurement.

2. Key Features

  • Ultra-low detection range: 0–10 ppm (customizable up to 5,000 ppm)

  • Fast response: T90 response time <10 seconds

  • High accuracy: ±3% Full Scale (F.S.)

  • Self-powered operation: No external power supply needed

  • Superior stability: Monthly drift <1.5%; linearity ±2.5% F.S.

  • Compact & lightweight: Miniature package for space-constrained applications

  • Anti-poisoning design: Resistant to common interfering gases (e.g., CO₂, N₂)

3. Technical Specifications

3.1 Electrical Parameters

ParameterValue
Working PrincipleElectrochemical (galvanic cell)
Output Current800–500 μA (in ambient air)
Measurement Range0–10 ppm (customizable: 0–5,000 ppm)
Response Time (T90)<10 s
Accuracy±3% F.S.
Linearity±2.5% F.S.
Monthly Drift<1.5%
Temperature Coefficient4.0%/°C
Electrical InterfaceGold-plated pins (pin 1: positive, pin 2: negative)

3.2 Mechanical Parameters

ParameterValue
Housing MaterialABS (Acrylonitrile Butadiene Styrene)
DimensionsΦ12 mm × 8 mm
Weight~3 g
Inlet Diameter1.5 mm

3.3 Environmental Parameters

ParameterValue
Operating Temperature0–50°C
Operating Humidity0–95% RH (non-condensing)
Operating PressureAtmospheric pressure ±15%
Expected Lifespan>18 months (in air)
Shelf Life9 months (from delivery date)

4. Application Scope

  • Portable Gas Analyzers: Handheld oxygen detectors, multi-gas analyzers

  • Medical Industry: Anesthesia delivery systems, respiratory monitors, incubators

  • Industrial Process Control: Nitrogen generator monitoring, semiconductor packaging, food packaging

  • Laboratory Equipment: Inert gas cabinets, glove boxes, gas chromatographs

  • Safety Monitoring: Confined space entry, pipeline leak detection, inerting system verification

5. Installation & Operation Guidelines

5.1 Installation

  1. Orientation: Mount horizontally or vertically; avoid direct exposure to high-velocity gas flow.

  2. Wiring: Connect pin 1 (positive) and pin 2 (negative) to the measurement circuit; avoid reverse polarity.

  3. Gas Connection: Use 1/16-inch PTFE tubing; maintain flow rate 100–300 mL/min (excessive flow may damage the sensor).

5.2 Operation

  1. Warm-Up: Allow 10-minute warm-up after installation for stable readings.

  2. Calibration:

    • Zero Calibration: Expose to 100% nitrogen (0 ppm O₂) and adjust zero offset.

    • Span Calibration: Expose to a known concentration (e.g., 10 ppm) and adjust span gain.

  3. Reading: The sensor outputs a current signal proportional to oxygen concentration; convert to ppm via the calibration curve.

6. Safety & Maintenance

6.1 Safety Precautions

  • Non-explosion-proof: Do not use in flammable/explosive atmospheres.

  • No liquid contact: Avoid exposure to water, oil, or corrosive gases (e.g., H₂S, Cl₂).

  • ESD protection: Use anti-static gloves during handling to prevent electrostatic damage.

6.2 Maintenance

  • Regular Calibration: Perform zero/span calibration every 3–6 months (depending on usage).

  • Filter Replacement: Replace the inlet filter every 4 months to prevent dust contamination.

  • Storage: Store in a cool, dry place (15–25°C, 40–60% RH) in sealed packaging; avoid extreme temperatures.

7. Troubleshooting

IssuePossible CauseSolution
No OutputWiring error or sensor expirationCheck wiring; replace sensor if expired
Unstable ReadingsInsufficient warm-up or flow turbulenceWarm up for 10 min; stabilize gas flow
Slow ResponseContaminated filter or sensor agingReplace filter; recalibrate or replace sensor
Large DriftTemperature fluctuation or expired calibrationCompensate for temperature; recalibrate

8. Manufacturer Information

  • Brand: ASAIR (Guangzhou Aosong Electronics Co., Ltd.)

  • Address: No. 10, Kexue Avenue, Science City, Guangzhou, China

  • Website: www.aosong.com

  • Support Email: support@aosong.com

  • Tel: +86-20-3229 8888

9. Quality Guarantee

ASAIR warrants the CM-2 to be free from defects in materials and workmanship for 12 months from the delivery date. This warranty excludes damage from misuse, improper installation, or harsh environmental exposure.