YOKOGAWA DS600-00-1W Product Description
November 14, 2026

YOKOGAWA DS600-00-1W Product Description

YOKOGAWA DS600-00-1W is a Yokogawa DA100 Darwin DS600 six-module data acquisition subunit for the DA100-2 expandable main unit, supplied at 100 to 240 V AC 50/60 Hz with W screw-terminal inlet variant, supporting up to six Yokogawa DA100 input, output or alarm modules as a replacement discontinued monitoring system spare.

Description

YOKOGAWA DS600-00-1W Product Description

Overview

YOKOGAWA DS600-00-1W is a Yokogawa DA100 Darwin six-module data acquisition subunit. In the DA100 expandable system, the DS600 acts as an interface between the DA100-2 main unit and up to six input, output or alarm modules. It provides the module mounting, subunit connection, power distribution and signal interface needed to collect measurement signals from Yokogawa DA100 input modules. The DS600-00-1W uses AC supply and the W power inlet or terminal variant. It is a replacement Darwin subunit spare for existing Yokogawa DA100 monitoring and recording installations.

Features

The DS600 is the six-module subunit in the DA100 product family, compared with the DS400 four-module subunit. It accepts up to six DA100 input or output modules such as universal DCV, thermocouple, RTD and digital input modules; DCV/TC/DI dedicated modules; power monitor; strain; pulse; mA; alarm contact output; or DI/DO modules, depending on the system configuration. The subunit does not itself generate a fixed channel count; available measurement channels depend on the installed input modules.

The suffix code for DS600-00-1W follows the DA100 subunit ordering structure. Supply suffix 1 indicates 100 to 240 V AC power, matching the official DA100 subunit AC supply specification. The final W indicates a 3-pin inlet with screw conversion terminal per the DA100 subunit power inlet code table. Distributor records for the DS600-00-1W list supply 100 to 240 V AC, frequency 50/60 Hz, and maximum power consumption 70 VA.

The DS600 dimensions and weight are listed in the DA100 specification with I/O modules installed: approximately 422 mm width × 176 mm height × 100 mm depth, and approximately 3.5 kg. The subunit connects to the DA100-2 main unit through the DA100 dedicated extension or subunit connection; extension cable type and length must match the original system layout.

Specs

Brand: YOKOGAWA, Yokogawa Test & Measurement / Yokogawa Electric Series: DA100 Darwin data acquisition system Product type: Subunit, module mounting and interface unit Model: DS600-00-1WSubunit family: DS600, six-module subunit Compatible main unit: DA100-2 expandable DA100 main unit Maximum module slots: 6Supported modules: Universal DCV/TC/RTD/DI, DCV/TC/DI dedicated, power monitor, strain, pulse, mA, alarm contact output, DI/DO and compatible DA100 modules Supply type: 100 to 240 V ACRated supply voltage: 100 to 240 V ACUsable AC supply: 90 to 250 V AC per DA100 general specification Rated frequency: 50/60 HzMaximum power consumption: 70 VA per distributor record Power inlet or cable suffix: W, 3-pin inlet with screw conversion terminal Dimensions with I/O modules: Approximately 422 W × 176 H × 100 D mmWeight with I/O modules: Approximately 3.5 kgInput channels: Determined by installed input modules, not fixed by the DS600 subunit Connection to main unit: DA100 dedicated subunit or extension connection; cable ordered separately Product status: Discontinued Yokogawa DA100 Darwin subunit; replacement spare for existing systems

Principle

The DA100 DS600 subunit mounts the input and output modules used to interface field sensors and process signals. Each installed input module conditions and digitizes its assigned signals, such as temperature from thermocouples or RTDs, DC voltage, mA current, strain bridge signals, pulse count, power measurement or contact inputs. The subunit transfers module data and control information between the connected DA100-2 main unit and the installed modules.

The main unit coordinates measurement timing, configuration, alarm handling and PC communication through its own communication module. The DS600 provides the physical module slots, power connection and signal interconnection for the measurement modules. External power supplies the subunit through the W inlet or terminal variant, while field wiring connects to the terminal or clamp terminals of the installed input modules rather than to the subunit itself.

Replacement should verify the DS600 model and suffix, connected DA100-2 main unit, subunit or extension cable type and length, installed module models and slot positions, module wiring, power supply inlet and terminal arrangement, system configuration or parameter backup, and communication settings. Because DA100 systems are discontinued, the replacement subunit should be checked for compatibility with the existing modules, main unit and legacy DAQ32 or DAQ32 Plus software before commissioning.

Applications

YOKOGAWA DS600-00-1W is used in legacy DA100 Darwin distributed data acquisition systems as a six-module expansion subunit for test benches, laboratory measurement, production monitoring, plant data logging, thermal process recording, machine condition monitoring, strain or load measurement, power monitoring, pulse counting, temperature distribution surveys, process variable recording and research or development testing. It supports installations requiring more input channels than a single DA100-2 main unit or one DS400 subunit can provide. In existing Darwin systems, the subunit serves as a direct replacement six-module spare matched to the original main unit, installed modules, cable connection and software configuration.