The 760C913A is a high-precision X-Ray Test Phantom & Calibration Standard designed for QA/QC of radiographic, fluoroscopic, and mammographic systems. It integrates multi-density targets, resolution grids, and geometric alignment markers to verify imaging performance.

Description

760C913A.jpg

1.1 Description
The 760C913A is a high-precision X-Ray Test Phantom & Calibration Standard designed for QA/QC of radiographic, fluoroscopic, and mammographic systems. It integrates multi-density targets, resolution grids, and geometric alignment markers to verify imaging performance.
1.2 Key Applications
  • Resolution verification (2–20 LP/mm)

  • Contrast sensitivity assessment

  • Beam uniformity & artifact inspection

  • Geometric accuracy & alignment checks

  • Mammography system calibration

2. Safety & Compliance

  • Warning: For professional use only by qualified medical physics/radiology staff.

  • Follow facility radiation safety protocols; avoid overexposure.

  • Comply with IEC 61223, FDA, and local regulatory standards.

  • CE marked, NIST-traceable calibrated.

3. Specifications

表格
ParameterValue
MaterialRadiopaque/radiolucent medical-grade polymer
Test PatternsLine pairs, bar targets, low-contrast inserts, alignment marks
Dimensions (L×W×H)22 cm × 18 cm × 3.0 cm
Weight~1.5 kg
Operating Temp15–35°C
Storage Temp-10–50°C
CompatibilityX-ray, fluoroscopy, mammography (2D)

4. Operating Instructions

4.1 Setup
  1. Place phantom centered on the patient table/cassette holder.

  2. Align reference marks with the X-ray central ray.

  3. Set exposure parameters (kVp, mAs, SID) per protocol.

4.2 Test Workflow
  1. Resolution: Expose line-pair grid; record smallest resolvable LP/mm.

  2. Contrast: Image low-contrast bars; identify thinnest visible bar.

  3. Uniformity: Inspect for uneven density, noise, or artifacts.

  4. Geometry: Verify alignment and dimensional accuracy.

5. Maintenance

  • Cleaning: Wipe with mild non-abrasive disinfectant; avoid harsh chemicals.

  • Storage: Keep in dry, dust-free case at room temperature.

  • Inspection: Check for cracks/scratches quarterly; replace if damaged.

  • Calibration: Annual recalibration recommended (NVLAP-accredited lab).

6. Warranty & Support

  • Warranty: 2-year limited warranty against defects.

  • Contact: Fluke Biomedical | Cleveland, OH | +1-440-498-2564 | www.flukebiomedical.com