NI SCXI-1122
Description
I. Product Introduction
NI SCXI-1122 is a 16-channel isolation relay multiplexer module of the SCXI series. It adopts the "16-channel relay selection → single-channel isolated amplifier + low-pass filter" architecture, specifically designed for multi-channel, low-rate, and high-isolation requirements for sensor signal conditioning. It can interface with voltages ranging from millivolts to ±250V, thermocouples, RTDs, strain bridges, 4-20mA currents, etc. It integrates isolation excitation, cold junction compensation, and bridge completion functions. It is discontinued (Legacy), but can be repaired or purchased as a second-hand / in stock.
II. Model Interpretation
SCXI: Instrument Signal Conditioning Expansion Series
1122: Basic model, 16-channel isolation multiplexer (single amplifier shared)
Part Number: 776572-22
III. Technical Parameters (Official, 25℃)
Channels and Isolation
Channels: 16 single-ended / differential isolation inputs, isolation between channels and ground
Working Isolation Voltage: 250Vrms continuous (between channels) / 480Vrms (to ground), CE certified double insulation, CAT II
Input Range (Software Uniform Setting): ±5mV to ±250V
Input Impedance: 1GΩ (powered on) / 100kΩ (powered off)
Common Mode Rejection Ratio CMRR: ≥100dB (DC to 60Hz)
Signal Conditioning
Amplifier: Single-channel isolated amplifier (shared by 16 channels)
Gain: Programmable by software, covering millivolt to high voltage signals
Filter: Software selectable 4Hz or 4kHz low-pass (all channels unified), 3-pole RC, NMRR at 60dB at 4Hz
Excitation: 1-channel isolated voltage / current excitation (3.333V ± 0.04%, 1.0mA ± 0.04%), with remote sensing
Sampling and Precision
Scan Rate: ≤100S/s for all channels when filtered at 4kHz; ≤1S/s for relay switching rate limitation when filtered at 4Hz; (all channels)
Sampling Mode: Switch between single channels multiple times (not suitable for high-speed polling)
Non-Linearity: ≤±0.01% FSR
Gain Error (Calibrated): ≤0.05% reading
Temperature Drift: ≤10ppm/℃
Cold Junction Compensation (CJC): Built-in, supports thermocouples
Bridge Completion: Supports strain bridge, RTD measurement
Electrical and Environmental
Power Supply: 5V ≤ 30mA; ±15V ≤ 200mA (backplane of the chassis)
Operating Temperature: 0 to 50℃ (standard commercial grade)
Storage Temperature: -40℃ to +85℃
Humidity: 5% to 90% without condensation
Size / Weight: 1U SCXI standard (30×173×203mm) / approximately 0.8kg
IV. Interface and Communication Configuration
Front Panel: 96-pin DIN C connector, connects to SCXI-1322 (16-channel isolation terminals, with CJC and excitation) or SCXI-1303 (general terminals)
Back Panel: 50-pin ribbon cable connector, connects to the SCXI chassis backplane bus, transmits data, timing, and power supply
Communication: No independent communication port, relies on the SCXI bus to communicate with the backend DAQ device (such as PCIe-6251), driven and configured parameters, and data reading by NI-DAQmx
Cabling: Supports floating ground / grounded signals, compatible with thermocouples, RTDs, strain gauges, 4-20mA transmitters, high-voltage DC
V. Core Functions
16-channel High Isolation Multiplexer: 250Vrms isolation between channels, single amplifier shared design, saves cost and chassis space, suitable for low-speed multi-channel acquisition. Ultra-wide input range: ±5mV to ±250V, covering from weak signals to industrial high voltage, no additional conditioning required, simplifying the system.
Integrated sensor excitation and compensation: built-in isolated voltage/current excitation, CJC, bridge completion, directly supporting thermocouples, RTDs, strain bridges, one card for multiple uses.
Dual-band low-pass filtering: 4Hz (anti-interference against industrial frequency) / 4kHz (preserving dynamic signals) software selectable, suitable for static / quasi-static measurements.
High anti-interference capability: channel isolation + high common-mode rejection ratio, completely eliminating ground loops, suitable for industrial strong interference environments.
Six. Application scenarios
Multi-channel temperature monitoring: 16 channels of thermocouples / RTD temperature measurement (such as heat treatment furnaces, pipelines, tanks), with SCXI-1322 terminals to achieve CJC and isolation.
Strain and pressure measurement: signal conditioning for strain gauges, weighing sensors, pressure transmitters (4-20mA), suitable for structural testing, industrial weighing.
High-voltage / floating ground signal acquisition: high-voltage DC, signals from devices with different ground potentials, isolation protection for DAQ and personnel safety.
Low-speed high-density acquisition system: laboratories, production lines, equipment condition monitoring, with many channels, low rate, high interference requirements.
Seven. Usage and maintenance instructions
Usage instructions
Installation: Power off, insert SCXI chassis, tighten fixing screws; supports hot swapping, can replace modules during operation.
Connection: Firmly connect isolation terminals, good insulation,严禁 mixing high and low voltage connections; use SCXI-1322 for thermocouples, enable CJC; strain bridges need excitation and bridge completion.
Configuration: NI-DAQmx sets input range, filtering bandwidth, excitation mode; scan rate strictly limited by filtering level (4kHz ≤ 100S/s, 4Hz ≤ 1S/s).
System: 68-pin SCSI cable connects SCXI chassis and DAQ; multi-chassis cascading sets a unique address to avoid conflicts.
Maintenance instructions
Environment: Keep dry, prevent condensation; away from strong electromagnetic sources, vibration sources.
Regular inspection: Check terminals, connectors, cables every quarter, for any looseness, oxidation, damage; clean surface dust.
Calibration: Once a year, focus on calibrating gain, linearity, CJC; for thermocouple / strain measurement scenarios, calibrate once every half year.
Fault handling: Data drift → calibration; signal jump → check grounding, shielding, terminal tightening; abnormal excitation → check excitation circuit and wiring.