PXIE-6363
Description
Model Definition
PXIe: PXI Express bus (Gen2 x4)
6363: X‑series DAQ, high‑density analog I/O
Key configuration: 32 AI, 4 AO, 48 DIO, 4 counters
Form factor: 1‑slot 3U PXIe
Key Specifications
Analog Input (AI)
Channels: 32 single‑ended / 16 differential
Resolution: 16‑bit
Max sampling rate: 2 MS/s (single‑channel), 1 MS/s (multi‑channel aggregate)
Input ranges: ±0.1 V to ±10 V (7 ranges)
Input impedance: 10 kΩ
Coupling: DC
Analog Output (AO)
Channels: 4
Resolution: 16‑bit
Update rate: 2.86 MS/s
Output range: ±5 V, ±10 V, or external reference
Output impedance: 0.2 Ω
Current drive: ±5 mA
Digital I/O (DIO)
Lines: 48 (3.3 V TTL/CMOS)
Hardware‑timed: 32 lines up to 10 MHz
Function: Pattern I/O, PWM, triggering
Counters/TimersNI
4 × 32‑bit counters
Uses: PWM, encoder, frequency, event counting
GeneralNI
Power: ~18 W
Operating temp: 0 °C to 55 °C
Calibration: Factory calibrated, 2‑year interval
Part number: 781056‑01
Interface & Synchronization
Host: PXIe Gen2 x4, 1.6 GB/s
Timing: NI‑STC3 engine, 10 ns resolution, 50 ppm accuracyNI
Triggering: Analog/digital trigger, PXI trigger bus, retriggerable tasksNI
Synchronization: Multi‑module sync via PXI star trigger, external 10 MHz REF
Core Features
High‑density mixed‑signal I/O: Combines 32 AI, 4 AO, 48 DIO for comprehensive system control.
16‑bit precision: Ensures accurate closed‑loop control and waveform generation.
Independent timing engines: Separate AI/AO/DIO clocks enable complex, asynchronous tasks.
Streaming capability: DMA for continuous data streaming to/from host.
NI‑DAQmx driver: Simplifies configuration, synchronization, and measurement tasksNI.
Applications
Hardware‑in‑the‑Loop (HIL): Real‑time simulation and control of mechatronic systems.
Automated Test Equipment (ATE): Functional test of sensors, actuators, and control circuits.
Industrial Control: High‑speed closed‑loop control for motion, process, and power systems.
Aerospace & Defense: Avionics signal simulation, sensor data acquisition.
Research & Lab: High‑speed data logging, precision waveform generation.
Usage Notes
Installation: Insert into 1 slot of a 3U PXIe chassis; no external power needed.
Warm‑up: 15 minutes for rated accuracy.
Wiring: Use shielded cables for analog signals; separate analog/digital grounds.
Protection: Avoid overvoltage (AI ±11 V max, AO ±25 V protected).
Maintenance: Keep connectors clean; self‑calibrate annually.
Get a Quote