PXI-2546
Description
II. Model Interpretation
PXI: PXI Hybrid Bus (compatible with PXIe Hybrid Slot)
NI 2546: Serial Number, representing 2.7 GHz, dual 4×1, electromechanical RF multiplexer
Alias: NI-2546, Dual 4x1 Mux, Dual SP4T RF Switch
Functional Positioning: 2 sets of independent 4-out-of-1 RF switches, electromechanical low loss + full terminal, parallel test expansion, preferred for medium-low frequency mass production
III. Technical Parameters Basic parameters
Topology: 2 independent groups of 4×1 (SP4T) multiplexers, Bank A (CH0A to CH3A → COMA), Bank B (CH0B to CH3B → COMB)
Relay type: Electromechanical relay (EMR)
Frequency range: DC to 2.7 GHz (–3 dB)
Characteristic impedance: 50 Ω (nominal)
Terminal configuration: All channels have built-in 50 Ω terminations (automatically connected when not selected)
RF performance (typical / guaranteed values)
Insertion loss:
≤1 GHz: <0.75 dB (typical <0.55 dB) NI
≤1.8 GHz: <1.0 dB (typical <0.8 dB) NI
≤2.7 GHz: <1.35 dB (typical <1.1 dB) NI
Isolation:
1 GHz: >50 dB NI
2.7 GHz: >40 dB NI
VSWR: <1.3 (within 2.7 GHz) NI
Maximum switching power: +24 dBm (continuous) NI
Terminal power: 1.5 W @25 °C (temperature rating reduced at high temperatures) NI
Electrical and dynamic
Maximum switching voltage: 30 VAC / 42 VDC NI
Maximum switching current: 0.5 A (per channel)
Switching time: typical 5 ms (electromechanical) NI
Relay lifespan: 1,000,000 times (typical), board-mounted counting tracking NI
DC path resistance: initial <0.25 Ω; end of life ≥1.0 Ω NI
Physics and environment
I/O connector: RF 10×SMA female connector (8 channels + 2 common terminals) NI
Size: 3U PXI single slot (approximately 100×160×20 mm) NI
Weight: approximately 255 g
Power consumption: 12 V ≈0.35 W NI
Operating temperature: 0 to 55 °C NI
Storage temperature: –40 to 70 °C NI
Four. Interface and Communication Configuration
Front-end interface
RF interface: 10×SMA female connector (gold-plated) NI
Bank A: CH0A, CH1A, CH2A, CH3A, COMA
Bank B: CH0B, CH1B, CH2B, CH3B, COMB
Trigger interface: no independent trigger (supports PXI backplane hardware trigger) NI
Bus and control
Bus interface: standard 3U PXI mixed bus (compatible with PXIe mixed slots) NI
Control software: compatible with NI- SWITCH, MAX, LabVIEW, TestStand, Switch Executive NI
Driver support: NI-DAQmx, NI-VISA, provides VI and API function libraries NI
Trigger synchronization: PXI hardware trigger, nanosecond-level synchronization, supports multi-module cascading NI
Configuration method: software to set channel mapping, scan sequence, relay count reading NI
Adaptation and expansion
Adaptation cable: SMA-SMA direct connection; optional SMA to BNC/SMB/MCX NI
Expansion capability: multi-module cascading, expand channel quantity, build large-scale RF switch system NI
Five. Core Functions
Dual 4×1 independent switches: 2 groups of completely independent 4-select-1 routing, can simultaneously control two RF signals, support parallel testing, double throughput NI.
DC to 2.7 GHz wideband coverage: covers Bluetooth, WiFi, ISM, 2G/3G, etc. mainstream medium and low-frequency RF frequency bands, suitable for consumer electronics and industrial RF testing NI. Mechanical and Electronic Relay Low Loss: Compared to solid-state relays, it has lower insertion loss and lower cost, making it suitable for mass production scenarios where loss sensitivity and moderate switching frequency are required (NI).
Full-channel 50 Ω Terminal: Unselected channels automatically connect to the load to prevent damage to the signal source due to large power reflection, improving system stability and measurement repeatability (NI).
Relay Life Counter: On-board counter, real-time statistics of switching times, predicting lifespan and reducing unplanned downtime (NI).
PXI Standard Integration: Single-slot 3U size, seamlessly integrated with PXI chassis, compatible with automated test systems, supporting hardware trigger synchronization (NI).
Six, Application Scenarios
Consumer Electronics RF Testing: Function verification and mass production testing of Bluetooth, WiFi, 2G/3G, NFC modules (NI).
Multi-channel RF Signal Switching: Channel expansion for signal sources, spectrum analyzers, and network analyzers, parallel polling testing for multiple DUTs (NI).
Mass Production ATE Testing: Cost-sensitive, medium-frequency, and medium switching frequency mass production lines, balancing performance and budget (NI).
Research and Development Laboratory: RF signal routing, instrument interconnection, solution verification, low-loss design ensures test accuracy (NI).
Semiconductor RF Chip Testing: Function verification, parameter testing, aging screening of medium-frequency RF ICs (NI).
Seven, Usage and Maintenance Instructions
Usage Points
Installation: Standard 3U PXI single slot, lock the panel, ensure reliable contact of the backplane; avoid strong vibration and electromagnetic interference (NI).
Connection: SMA connectors should be gently inserted and removed, avoid violent operation; RF power ≤ +24 dBm,严禁 operating beyond power limit (NI).
Power-on State: By default, all channels are disconnected from COM; unselected channels automatically connect to 50 Ω terminal (NI).
Software Control: Use NI-SWITCH or LabVIEW VI for Connect/Disconnect; regularly read relay counter (NI).
Environment: Working temperature 0-55 °C, dust-proof, moisture-proof, and dew-proof; keep away from strong electromagnetic interference sources, ensure good grounding (NI).
Maintenance Instructions
Regular Inspection: Check SMA connectors for looseness, oxidation, or damage every quarter; clean the panel and ventilation ports, maintain good heat dissipation (NI).
Life Management: Mechanical and electronic relays have approximately 1 million switching times; regularly read the onboard counter, plan for replacement in advance to avoid downtime (NI).
Fault Troubleshooting: First check SMA connection and grounding for abnormal signals; use MAX to diagnose switch status; take precautions against static electricity to avoid damage to modules (NI).
Storage: Power off when not in use, cover SMA connector with dust cap, store dry and dust-free, avoid high temperature, high humidity, and strong magnetic field environments (NI).
Calibration: It is recommended to calibrate once a year to ensure that insertion loss, isolation, VSWR, etc. meet specifications; contact NI or a third-party calibration institution (NI).
Get a Quote