PXI-2546
May 27, 2026

PXI-2546

I. Product Introduction PXI-2546 is a 3U PXI dual 4×1 (dual SP4T) electromechanical RF multiplexer launched by National Instruments (NI). It is in production (Active), with part number 778572-46NI. The module uses electromechanical relays (EMR) and supports DC to 2.7 GHz, 50 Ω RF signal routing. It features dual independent channel groups and full-channel terminals, targeting medium and low frequency, parallel testing, and high-stability mass production RF ATE systems, while balancing low loss and cost advantages. NI.

Description

II. Model Interpretation

PXI: PXI Hybrid Bus (compatible with PXIe Hybrid Slot)

NI 2546: Serial Number, representing 2.7 GHz, dual 4×1, electromechanical RF multiplexer

Alias: NI-2546, Dual 4x1 Mux, Dual SP4T RF Switch

Functional Positioning: 2 sets of independent 4-out-of-1 RF switches, electromechanical low loss + full terminal, parallel test expansion, preferred for medium-low frequency mass production

III. Technical Parameters Basic parameters

Topology: 2 independent groups of 4×1 (SP4T) multiplexers, Bank A (CH0A to CH3A → COMA), Bank B (CH0B to CH3B → COMB)

Relay type: Electromechanical relay (EMR)

Frequency range: DC to 2.7 GHz (–3 dB)

Characteristic impedance: 50 Ω (nominal)

Terminal configuration: All channels have built-in 50 Ω terminations (automatically connected when not selected)

RF performance (typical / guaranteed values)

Insertion loss:

≤1 GHz: <0.75 dB (typical <0.55 dB) NI

≤1.8 GHz: <1.0 dB (typical <0.8 dB) NI

≤2.7 GHz: <1.35 dB (typical <1.1 dB) NI

Isolation:

1 GHz: >50 dB NI

2.7 GHz: >40 dB NI

VSWR: <1.3 (within 2.7 GHz) NI

Maximum switching power: +24 dBm (continuous) NI

Terminal power: 1.5 W @25 °C (temperature rating reduced at high temperatures) NI

Electrical and dynamic

Maximum switching voltage: 30 VAC / 42 VDC NI

Maximum switching current: 0.5 A (per channel)

Switching time: typical 5 ms (electromechanical) NI

Relay lifespan: 1,000,000 times (typical), board-mounted counting tracking NI

DC path resistance: initial <0.25 Ω; end of life ≥1.0 Ω NI

Physics and environment

I/O connector: RF 10×SMA female connector (8 channels + 2 common terminals) NI

Size: 3U PXI single slot (approximately 100×160×20 mm) NI

Weight: approximately 255 g

Power consumption: 12 V ≈0.35 W NI

Operating temperature: 0 to 55 °C NI

Storage temperature: –40 to 70 °C NI

Four. Interface and Communication Configuration

Front-end interface

RF interface: 10×SMA female connector (gold-plated) NI

Bank A: CH0A, CH1A, CH2A, CH3A, COMA

Bank B: CH0B, CH1B, CH2B, CH3B, COMB

Trigger interface: no independent trigger (supports PXI backplane hardware trigger) NI

Bus and control

Bus interface: standard 3U PXI mixed bus (compatible with PXIe mixed slots) NI

Control software: compatible with NI- SWITCH, MAX, LabVIEW, TestStand, Switch Executive NI

Driver support: NI-DAQmx, NI-VISA, provides VI and API function libraries NI

Trigger synchronization: PXI hardware trigger, nanosecond-level synchronization, supports multi-module cascading NI

Configuration method: software to set channel mapping, scan sequence, relay count reading NI

Adaptation and expansion

Adaptation cable: SMA-SMA direct connection; optional SMA to BNC/SMB/MCX NI

Expansion capability: multi-module cascading, expand channel quantity, build large-scale RF switch system NI

Five. Core Functions

Dual 4×1 independent switches: 2 groups of completely independent 4-select-1 routing, can simultaneously control two RF signals, support parallel testing, double throughput NI.

DC to 2.7 GHz wideband coverage: covers Bluetooth, WiFi, ISM, 2G/3G, etc. mainstream medium and low-frequency RF frequency bands, suitable for consumer electronics and industrial RF testing NI. Mechanical and Electronic Relay Low Loss: Compared to solid-state relays, it has lower insertion loss and lower cost, making it suitable for mass production scenarios where loss sensitivity and moderate switching frequency are required (NI).

Full-channel 50 Ω Terminal: Unselected channels automatically connect to the load to prevent damage to the signal source due to large power reflection, improving system stability and measurement repeatability (NI).

Relay Life Counter: On-board counter, real-time statistics of switching times, predicting lifespan and reducing unplanned downtime (NI).

PXI Standard Integration: Single-slot 3U size, seamlessly integrated with PXI chassis, compatible with automated test systems, supporting hardware trigger synchronization (NI).

Six, Application Scenarios

Consumer Electronics RF Testing: Function verification and mass production testing of Bluetooth, WiFi, 2G/3G, NFC modules (NI).

Multi-channel RF Signal Switching: Channel expansion for signal sources, spectrum analyzers, and network analyzers, parallel polling testing for multiple DUTs (NI).

Mass Production ATE Testing: Cost-sensitive, medium-frequency, and medium switching frequency mass production lines, balancing performance and budget (NI).

Research and Development Laboratory: RF signal routing, instrument interconnection, solution verification, low-loss design ensures test accuracy (NI).

Semiconductor RF Chip Testing: Function verification, parameter testing, aging screening of medium-frequency RF ICs (NI).

Seven, Usage and Maintenance Instructions

Usage Points

Installation: Standard 3U PXI single slot, lock the panel, ensure reliable contact of the backplane; avoid strong vibration and electromagnetic interference (NI).

Connection: SMA connectors should be gently inserted and removed, avoid violent operation; RF power ≤ +24 dBm,严禁 operating beyond power limit (NI).

Power-on State: By default, all channels are disconnected from COM; unselected channels automatically connect to 50 Ω terminal (NI).

Software Control: Use NI-SWITCH or LabVIEW VI for Connect/Disconnect; regularly read relay counter (NI).

Environment: Working temperature 0-55 °C, dust-proof, moisture-proof, and dew-proof; keep away from strong electromagnetic interference sources, ensure good grounding (NI).

Maintenance Instructions

Regular Inspection: Check SMA connectors for looseness, oxidation, or damage every quarter; clean the panel and ventilation ports, maintain good heat dissipation (NI).

Life Management: Mechanical and electronic relays have approximately 1 million switching times; regularly read the onboard counter, plan for replacement in advance to avoid downtime (NI).

Fault Troubleshooting: First check SMA connection and grounding for abnormal signals; use MAX to diagnose switch status; take precautions against static electricity to avoid damage to modules (NI).

Storage: Power off when not in use, cover SMA connector with dust cap, store dry and dust-free, avoid high temperature, high humidity, and strong magnetic field environments (NI).

Calibration: It is recommended to calibrate once a year to ensure that insertion loss, isolation, VSWR, etc. meet specifications; contact NI or a third-party calibration institution (NI).


Get a Quote