PXI-2503
Description
II. Model Interpretation
PXI: PCI eXtensions for Instrumentation, a PCI expansion bus standard for instruments.
2503: 2500 series (PXI switch module), 03 represents 48 channels, electromechanical relays, 1/2/4 lines, medium voltage type.
Part Number: 777697-01 (standard model).
Status: In stock (NI's current mainstream switch module).
III. Technical Parameters
Channel Configuration (Software Configurable):
1 line: 48×1 multiplexer
2 lines: 24×1, dual 12×1, four 6×1 multiplexers
4 lines: 12×1 multiplexer
Matrix: 2 lines 4×6 matrix
Electrical Characteristics:
Maximum switching voltage: 60 VDC, 30 VAC (effective value)
Maximum switching/carrying current: 1 A (per channel)
Maximum switching power: 30 W (per channel)
Relay Type: Electromechanical, low thermal offset (<1 μV)
Conduction Resistance: <1.5 Ω (typical)
Dynamic Performance:
Bandwidth: ≥10 MHz (50 Ω termination)
Scan Rate: Up to 100 channels/second
Scan List: 1,024 steps (deterministic scan)
Relay Life: ≥10^6 times (rated load)
Physical and Environmental:
Size: 3U PXI (100 mm×160 mm)
Weight: Approximately 0.23 kg
Operating Temperature: 0–50 °C; Storage: −20–70 °C
Humidity: 10%–90% RH (no condensation)
Power Consumption: Typical 370 mA, maximum 700 mA (+5 V) Data acquisition channel expansion: Expansion of PXI DAQ module channels to achieve synchronous high-channel number acquisition.
Research and laboratory: Material testing, electrochemical measurement, multi-channel monitoring, and precise instrument calibration by NI.
VII. Installation and Maintenance Instructions
Installation and Wiring:
Module: Unplug and then plug in, insert into the 3U slot of the PXI chassis, and tighten the fixing screws.
Terminal block: Use TB-2605, connect according to the topology (1/2/4 wires), avoid short circuits.
Grounding: Reliable connection of signal ground to chassis ground to reduce noise.
Software Configuration:
Install NI-SWITCH driver, NI-MAX recognizes the module, configure topology and scan list by NI.
Programming: Directly call the switch VI in LabVIEW, or control channel switching and scanning using C#/Python.
Daily Maintenance:
Cleaning: After power off, wipe the panel with a soft cloth to avoid liquid entering the interface.
Environment: Working temperature 0–50 °C, avoid dust, moisture, and strong electromagnetic interference.
Relay lifespan: Avoid overpressure / overcurrent (>60 V/1 A), regularly check the contact status.
Static protection: Wear an anti-static wristband when plugging in and out the module, ensure good grounding of the chassis.
Get a Quote