330851-04-000-020-10-00-05
Description
Model Explanation
330851: Special probe for 25mm differential expansion
04: Standard M30×2 threaded housing
000: Standard size without threaded section
020: Total system length 2.0 meters
10: High-temperature non-armored cable
00: Tail-out structure
05: ATEX/IECEx explosion-proof certification
Core Parameters
Electrical Parameters
Sensitivity: 0.787V/mm
Linear range: 0~12.7mm
Frequency response range: 0~5kHz
Output: Negative pressure gap voltage signal
Mechanical Specifications
Installation thread: M30×2
Outlet method: Tail-out
Cable length: 2m
Minimum bending radius: ≥50mm
Protection level: IP67
Environmental Parameters
Probe short-term temperature resistance: 150℃
Cable operating temperature: -40℃~120℃
Vibration resistance level: 10g
Explosion-proof level: Ex ia IIC T4
System Compatibility
Matching preamplifier: 330850-020-05
Matching extension cable: 330854-020-24-05
Compatible rack: 3300 TSI, 3500 TSI
Compatible module: 3500/45M differential expansion displacement module
Installation Requirements
The length specification of the entire system is unified
Cables must not be compressed or bent with sharp corners
The shielding layer is grounded only at one end on the preamplifier side
Explosion-proof areas should be installed in accordance with the specifications Applicable scenarios
Monitoring of shaft expansion difference and shaft displacement for turbine generator sets
High-temperature factory area, hydrogen generation unit, chemical explosion-proof area
Scenarios of point installation over short distances
Get a Quote