PCI4462
Description
I. Basic Information
Product Name: PCI-4462
Part Number: 779308-01 779309-01
Manufacturer: National Instruments NI
Product Type: PCI bus high-precision dynamic signal acquisition card DSA
Status: Out of production as of 2026, no new stock available
Alternative Model: PXIe-4462
Features: 4 channels of input, no output, 24-bit, 118dB, multi-channel IEPE
II. Main Applications
Acoustic testing, microphone array, vibration testing, NVH analysis, modal experiment, noise analysis, industrial equipment fault diagnosis, multi-channel synchronous acquisition
III. Analog Input
Number of channels: 4 differential synchronous sampling channels
Resolution: 24-bit Delta-Sigma ADC
Sampling rate: 1kS/s to 204.8kS/s programmable
Analog bandwidth: DC to 90kHz
Dynamic range: 118dB
Total harmonic distortion: -100dB at 1kHz
Input coupling: Software selectable AC/DC per channel independent
IEPE conditioning: Software selectable 0mA/4mA/10mA per channel independent
Input range: 6 gain levels programmable
Gain: -12dB ±42.4V, -6dB ±31.6V, 0dB ±10V, 6dB ±3.16V, 12dB ±1V, 18dB ±0.316V
Input impedance: 1MΩ, parallel 240pF (IEPE enabled)
IV. Analog Output
No analog output
V. Digital IO and Trigger
Digital trigger: 1 SMB external trigger
No general digital IO
No counter
VI. Synchronization and Trigger
Trigger mode: Software trigger, hardware external trigger
Synchronization bus: Supports RTSI bus, multi-card synchronization
Anti-aliasing: Built-in high-order programmable anti-aliasing filter
TEDS: Supports sensor electronic data sheet IEEE1451
VII. Electrical and Mechanical
Bus interface: PCI 32-bit 5V/3.3V compatible
Board size: Standard PCI half-length card
Power supply: PCI slot power supply
Operating temperature: 0 to 55 degrees Celsius
Relative humidity: 10% to 90% without condensation
Connector: BNC, analog input SMB trigger
Weight: Approximately 227 grams
VIII. Software Support
Driver: NI-DAQmx
Development environment: LabVIEW, LabWindowsCVI, Visual Studio
Operating system: Windows XP, 7, 10, 11 32/64-bit
IX. Key Features
24-bit ultra-high precision, dynamic range 118dB
4 AI channels synchronous acquisition, no AO suitable for multi-channel sensing
Each channel independent IEPE 0/4/10mA selectable
6 gain levels input range flexible adaptation
Built-in anti-aliasing filter, TEDS sensor identification
RTSI multi-card synchronization, BNC interface plug-and-play
Suitable for high-precision multi-channel audio acoustic vibration measurement

Get a Quote