SCXI-1102B | National Instruments(NI) 32-Ch Differential Voltage Input Conditioning Module
Description
Part Code Definition
SCXI = SCXI signal conditioning platform
1102B = revised version of SCXI-1102, fixed filter at 200Hz cutoff frequency, upgraded input protection circuit vs early SCXI-1102(2Hz filter option).
Technical Specifications
Input channel:32 fully differential input, DC-coupled
Full-scale range:±10V(Gain×1) / ±100mV(Gain×100); programmable gain ×1/×2/×5/×10/×20/×50/×100 software selectable
Per-channel filter:fixed 3rd-order low-pass filter, 200Hz cutoff frequency
Max sampling rate:333kS/s full-chassis total scan speed
Input impedance:1MΩ//12pF differential input; CMRR>110dB@50Hz
Overvoltage protection:each channel withstand ±42VAC peak /60VDC max transient input protection
Operation environment:0℃~+50℃ working, -40℃~+70℃ storage, RH5%~95% non-condensing
Dimension:standard SCXI module 173×86×203mm, net weight≈1.38kg
Certification:CE, UL, C/US certified
Interface & Communication Configuration
Front terminal:96-pin DIN connector, match SCXI-1302/SCXI-1303 screw terminal blocks for field sensor wiring; terminal embedded CJC temperature sensor for thermocouple cold end compensationNI
Rear port:SCXI standard backplane golden finger, plug into SCXI-1000/SCXI-1001 chassis slot, transmit multiplexed analog signal via chassis internal bus to PCI/PCIe DAQ card
Bus mode:32 channels time-division multiplex onto single SCXI analog bus to DAQ input channel
Core Functions
Multiplex 32 independent differential sensor signals into single programmable instrumentation amplifier circuit for DAQ channel expansion
Software adjustable PGA amplifies weak thermocouple mV signal and standard voltage/current signal uniformly
Fixed per-channel 200Hz low-pass filtering eliminates industrial power frequency and high-frequency noise interference
Onboard CJC cold-junction compensation realizes direct thermocouple temperature acquisition without extra compensation module
Complete overvoltage safeguard prevents field wiring surge from damaging module inner circuit
Application Scenarios
Automotive bench multi-point temperature & strain testing, aerospace environmental chamber multi-sensor centralized acquisition, petrochemical field process analog signal collection, lab thermal cycling test system, old SCXI DAQ system spare replacement.
Operation & Maintenance Notes
Power off SCXI chassis before plug/unplug module to avoid backplane burnout; separate weak signal twisted pair and high-power AC cable to reduce EMI interference; set gain parameter matching sensor output range in LabVIEW/NI-DAQmx before commissioning; regularly fasten terminal block screws to prevent contact drift; CJC sensor wiring strictly follow terminal definition to guarantee accurate thermocouple reading.