SCXI-1100 | National Instruments(NI) 32-Ch Differential Analog Input Conditioning Module
Description
Part Code Definition
SCXI = Signal Conditioning Extension for Instrumentation series platform
1100 = economical 32-channel multiplexed analog input signal conditioning module
Technical Specifications
Input channel:32 differential input, support single-ended wiring conversion via terminal block
Input range:DC coupled, full scale ±10V max input
Programmable gain:×1/×2/×5/×10/×20/×50/×100/×200/×500/×1000 software selectable
Input impedance:100MΩ per channel; CMRR 120dB @50Hz
Filter option:jumper configurable 4Hz low-pass anti-aliasing filter, optional higher frequency filter via terminal wiring
Sampling rate:Max 166kS/s full bandwidth;6.6kS/s@10kHz filter;3S/s@4Hz filter
Operating environment:0℃~+50℃ working, -40℃~+70℃ storage;RH5%~90% non-condensing
Dimension:173×86×203mm,net weight≈1.36kg
Certification:CE,UL,C/US safety certified
Interface & Communication Configuration
Front connector:96-pin DIN-C male socket, match SCXI-1300/SCXI-1303/SCXI-1308 series screw terminal blocks for field sensor wiring; terminal integrates cold-junction compensation CJC temperature sensor for thermocouple measurement
Rear connector:50-pin ribbon cable port, connect SCXI chassis backplane bus, cascade to PCI/PCIe DAQ acquisition board via chassis rear bus line
Bus mode:All 32 channels time-division multiplex to single DAQ input channel via SCXI backplane analog bus
Core Functions
Multiplex 32 independent differential sensor signals into single PGA circuit, realize high-density channel expansion for limited-channel DAQ card
Software programmable gain amplifying adapt weak mV thermocouple/strain gauge signal and standard volt sensor signal
Hardware jumper low-pass filtering suppress industrial power frequency interference, improve weak signal measuring stability
Built-in CJC cold-junction compensation circuit cooperated with matched terminal block to finish thermocouple temperature acquisition without extra compensation module
Compatible with NI-DAQmx,LabVIEW,LabWindows/CVI software for channel parameter configuration, real-time data reading and logging
Application Scenarios
Multi-point temperature monitoring of production line, lab environmental test bench multi-sensor acquisition, old SCXI DAQ system spare replacement, mechanical fatigue test multi-channel strain signal collection, petrochemical site field process analog signal centralized conditioning.
Operation & Maintenance Notes
Install into standard SCXI-1000(4-slot)/SCXI-1001(12-slot) chassis slot, power off chassis before plug/unplug module to avoid backplane damage; separate weak signal twisted wire and high-power AC cable to reduce EMI; configure gain and filter mode via LabVIEW before field commissioning; regular check terminal screw tightness preventing loose contact drift; CJC sensor wiring strictly follow terminal definition for accurate thermocouple reading.