350500-010000-01-01
Description
I. Model Code (350500-XX-XXXX-XX-XX)
According to the standard 350500 structure:
350500: Dynamic Pressure Charge Amplifier (DPCA)
01: Input Sensitivity = 8 pC/psi (116 pC/bar)
0000: Filter/Gain Combination = Standard (Low Pass + Standard Gain)
01: Output Sensitivity = 100 mV/psi (1.45 V/bar)
01: Installation/Connection = DIN Rail, MIL Connector
II. Core Parameters
1. Input (Compatible with High Temperature Pressure Transducer)
Input Sensitivity: 8 pC/psi (116 pC/bar)
Input Impedance: ≥ 10¹⁴ Ω
Maximum Input Charge: ±5000 pC
Sensor Type: High Temperature Piezoelectric (e.g. 330400 Series)
2. Output
Output Sensitivity: 100 mV/psi (1.45 V/bar)
Output Impedance: < 100 Ω
Frequency Response: 2 Hz ~ 10 kHz (-3dB)
Range: 0–50 psi pk (0–3.45 bar pk)
Accuracy: ±2% @ 150 Hz
3. Electrical
Power Supply: –24 VDC (±10%), approximately 20 mA
Power Consumption: ≤ 0.5 W
Interface: Compatible with Bently Eddy Current Preamplifier (–24V, Common, Signal)
4. Mechanical and Environmental
Size: 88 × 36 × 70 mm
Weight: Approximately 0.24 kg
Installation: DIN Rail
Operating Temperature: 0°C ~ +65°C
Storage Temperature: –40°C ~ +85°C
Humidity: 5%–95% RH (No Condensation)
Protection: IP65 (Front Panel)
III. Typical Accessories
Sensor: 330400 Series High Temperature Piezoelectric Pressure Sensor
Monitor: 3500/64 Dynamic Pressure Monitor
Extension Cable: Low Noise Special Cable (5m/9m)
IV. Key Features
High Temperature Compatibility: Designed for high-temperature environments of gas turbines
Low Noise: Low-noise design of the charge amplification stage
Standard Interface: Directly compatible with 3500 system, no conversion required
Long-term Stability: Drift < 0.1%/ Year, suitable for long-term online monitoring
Get a Quote