Model T8019 is designed per GB/T 8019 and ASTM D381 standards, for determining existent gum content in motor gasoline, aviation gasoline, and volatile distillates (not for aviation turbine fuels). It features precise temperature control, adjustable air jets, and simultaneous multi-sample testing.

Description

T8019.jpg

Model T8019 is designed per GB/T 8019 and ASTM D381 standards, for determining existent gum content in motor gasoline, aviation gasoline, and volatile distillates (not for aviation turbine fuels). It features precise temperature control, adjustable air jets, and simultaneous multi-sample testing.

2. Technical Specifications

ParameterValue
Power SupplyAC 220V±10%, 50Hz
Max Power≤2.5kW
Evaporation BathAluminum block, φ260×130mm
Test Holes3 (φ51×70mm each)
Working Temp160–165°C (±0.1°C accuracy)
Air Pressure0.035–0.07MPa (adjustable)
Air Flow1000±100mL/min per jet
Ambient Conditions0–40°C, ≤85% RH (non-condensing)
Dimensions550×450×350mm

3. Key Features

  • Standard Compliance: Meets GB/T 8019 and ASTM D381 for accurate gum content analysis.

  • Dual Control: Digital temperature controller (±0.1°C) and independent flowmeters for each jet.

  • Efficient Testing: 3 simultaneous test positions with uniform heat distribution.

  • Safe Design: Over-temperature protection, stainless steel bench, and oil-water separator.

  • Easy Operation: Desktop structure, intuitive controls, and built-in silent air compressor.

4. Operating Instructions

  1. Preparation: Place on a horizontal surface, connect power and air supply. Ground the unit securely.

  2. Warm-up: Turn on the heater; wait for the bath to stabilize at 160–165°C.

  3. Sample Loading: Pipette 50mL of cooled sample into a pre-weighed beaker (φ51×70mm). Insert into a test hole.

  4. Air Adjustment: Set air pressure to 0.035MPa; adjust each jet flow to 1000±100mL/min.

  5. Test Run: Start the timer (30min for gasoline). The instrument maintains temperature and airflow automatically.

  6. Completion: After 30min, stop airflow. Remove beakers, cool in a desiccator, and weigh to calculate gum content.

5. Safety Warnings

  • High Temperature: Avoid contact with the 160°C bath during operation—risk of burns.

  • Ventilation: Use in a fume hood or well-ventilated area to prevent vapor accumulation.

  • No LPG: Do not test liquefied petroleum gas—risk of overpressure.

  • Disconnect Power: Unplug before cleaning, maintenance, or moving the unit.

  • Grounding: Ensure proper grounding to prevent electric shock.

6. Maintenance

  • Daily: Clean beakers and jets after use; check air filters for contamination.

  • Weekly: Verify temperature calibration with a precision thermometer; adjust if needed.

  • Monthly: Drain and clean the oil-water separator; inspect hoses for leaks.

  • Storage: Keep in a dry, dust-free environment; cover to prevent contamination.