3051C*6538276
Description
1. Product Overview
2. Core Technical Parameters
Basic Accuracy: Reference accuracy up to ±0.04% FS; total performance error ≤±0.15% full span
Turndown Ratio: Max 150:1 adjustable measuring range
Output Signal: Standard 4–20mA DC + HART 5 / HART7 digital protocol; optional FOUNDATION Fieldbus / Profibus PA versionsEmerson
Supply Voltage: 10.5~42.4VDC for 4–20mA loop-powered operation
Long-term Stability: 5-year stability ≤±0.125% of calibrated range; up to 10-year long-term stability design
Process Wetted Parts: Standard 316L stainless steel; optional Hastelloy C, Monel, Tantalum for corrosive mediumEmerson
3. Function & Diagnostic Features
Optional local backlit LCD display for on-site real-time reading and zero/span trim
Built-in advanced diagnostics: plugged impulse line detection, loop circuit fault monitoring, process abnormal alert to reduce unexpected shutdown
Optional local push buttons for on-site calibration without external tools
Available SIL2/3 safety certification per IEC61508, FM/ATEX explosion-proof approvals for hazardous area installationEmerson
4. Ambient Working Condition
Operating ambient temperature: −40℃ ~ +85℃
Relative humidity: 5%~95% non-condensing
Process medium temperature: dependent on sensor fill fluid option, standard −40℃ ~ +120℃
5. Application Range
DP flow measurement with orifice/annubar flow elements
Tank liquid level measurement via differential pressure
Pipeline process pressure monitoring for gas & liquid medium
Power plant boiler feedwater, chemical reactor pressure control
Get a Quote